Dichtemessgerät Excellence D4
D4 Excellence density meters offer particularly high measuring speeds with automatic temperature regulation in the range from 0 to 95 °C and a precision of 0.0001 g/cm³. The One Click™ user interface enables users to start a complete workflow with one click including sampling, measuring, rinsing, drying, and SOPs. Measurement limits can be set for color-coded pass/fail information for accepted and rejected samples. Measurement limit values can be set for the color coding of accepted and rejected samples. • Automatic Temperature Control • User management • Status Light : The status light of the instrument shows if it is ready for a measurement (green), busy (blinking), or waiting for operator action (orange). • Automated Sampling, Rinsing and Drying • Built-in Measurement Scales: Density / Specific gravity, Alcohol tables, Acids / bases and chemical tables, User programmable tables / functions, Brix / Sugar tables • Integrated drying pump • Video view • Automatic sample detection • Automatic bubble detection (BubbleCheck™) • Automatic viscosity correction • Connection option for barcode readers • 7" display with color touchscreen • LIMS connectivity • Compliance/Data Security: Audit Trail (requires LabX software), Data Integrity (requires LabX software), 21 CFR Part 11/EU Annex 11 (requires LabX software) • Connectivity: HDMI (Terminal), USB client (PC Software), Ethernet (Network and PC Software), 3 × USB host (printer, memory stick, barcode reader, hub, keyboard, mouse), RS232 (Automation units), CAN (Multiple LevelSens) • Languages: English, Italian, French, Russian, Polish, Portuguese, Chinese, Korean, German, Spanish • Dimensions (W x D x H): 267 x 256 x 226 mm • Weight: 17.5kg Accuracy valid under ideal conditions and for samples with low densities and viscosities.
Article nr. | Name | PackageContent | Genauigkeit | Messbereich | Typ | Probenvolumen | Auflösung | Wiederholbarkeit | Normen und Standards | Temperatursteuerbereich |
---|---|---|---|---|---|---|---|---|---|---|
30474900 | 1 | ± 0,0001 g/cm3 | 0 - 3 g/cm3 | Excellence D4 | 1,50 ml | 0,0001 g/cm3 | ± 0,00005 g/cm3 | Ph. Eur. 2.2.5, DIN EN ISO 3675, ISO 2811-3, AOAC Offizielle Methode 988.06, IP 200/52, GBT 11540–2008, ASTM D 4806, API Std 2540, ASTM D 1475, USP Kapitel 841 (ab USP 29), GB/T 2013-2010, ISO 18301, EN 14214:2003, ASTM D 2501, ASTM D 6448, ASTM D 1250, DIN 51757, Chinesische Pharmakopöe 0601, GB/T 13531.4-2013, ISO/EN 15212-1, ISO 12185-2, Pharm. Eu. (2005) JP (2005) USP (2006), IP 365, ASTM D 5002, ISO 91, ANSI 1250, GB5009.225-2016, ASTM D 4052, OIV-Analysemethode zur Dichtemessung in Wein und Most, ASTM D5931, OIV MA-BS-03 und/oder MA-BS-03, JJF1070-2005, | 0 – 95 °C |